An intact mass confirms a peptide's composition. To confirm its sequence, the order of the residues, the peptide is broken apart inside the mass spectrometer and the pieces are weighed. This is tandem mass spectrometry (MS/MS). This article explains how peptides fragment, how a fragment ladder is read, which fragmentation methods suit which questions, and what MS/MS cannot settle.
Principle
In the first stage, the instrument selects one ion, the precursor, usually a protonated form of the peptide such as [M+2H]2+. It then gives that ion enough energy to break, and the second stage measures the masses of the fragments. Because peptides tend to break along the backbone, between residues, the fragments form series whose masses step up one residue at a time.
Fragment ion types
Breaking the backbone splits the peptide in two. By the nomenclature of Roepstorff and Fohlman, later refined by Biemann, fragments that keep the N-terminus are called a, b or c ions, and fragments that keep the C-terminus x, y or z ions, depending on which backbone bond broke. The number gives how many residues the fragment contains.
| Method | How energy is added | Main ions | Good for |
|---|---|---|---|
| CID (collision-induced dissociation) | Collisions with an inert gas | b and y | Routine sequencing of small, protonated peptides |
| HCD (higher-energy collisional dissociation) | Collisions at higher energy, in a separate cell | b and y, more y; low-mass immonium ions | High-resolution fragment spectra; marker ions |
| ETD (electron-transfer dissociation) | An electron transferred from a reagent anion to a multiply charged precursor | c and z | Longer, highly charged peptides; keeping labile modifications attached |
Reading a ladder
The difference between two neighboring ions in a series equals the mass of one residue, which identifies it: 57.021 Da for glycine, 71.037 for alanine, 113.084 for leucine or isoleucine, and so on. Reading the b series from low to high mass gives the sequence from the N-terminus; reading the y series gives it from the C-terminus. The two series run in opposite directions and confirm each other.
A worked example: for KPV, the b ions are at 129.102 (Lys) and 226.155 (Lys-Pro), and the y ions at 118.086 (Val) and 215.139 (Pro-Val). The step from 129.102 to 226.155 is 97.053 Da, a proline. Every peptide reference entry lists its expected b and y ions, and the peptide calculator gives them for any linear sequence.
Real spectra are not complete ladders. Some bonds break readily (the bond before proline is a well-known example), others rarely, and ions lose water or ammonia. A sequence is confirmed when enough consecutive steps are seen in at least one series, ideally both.
De novo sequencing and database matching
For a known synthetic peptide the question is confirmation: do the observed fragments match those predicted for the stated sequence? That is simple and robust. Reading an unknown sequence from scratch, de novo sequencing, is harder, because gaps and ambiguous steps leave several candidate sequences. Proteomics avoids the problem by matching spectra against sequence databases, an approach that suits natural proteins rather than modified synthetic peptides.
What MS/MS cannot settle
- Leucine and isoleucine have the same residue mass, so a ladder step cannot tell them apart without specialized fragmentation.
- Stereochemistry. A D-residue gives the same fragments as its L-form. Chiral analysis is needed.
- Rings. In a peptide closed by a disulfide or a lactam bridge, a single backbone cleavage inside the ring does not release a fragment. Reduction before analysis, or more complex interpretation, is needed.
- Quantity. Fragment spectra confirm identity; they do not measure purity.
Frequently asked questions
- Why is the doubly charged ion usually chosen as the precursor?
- In electrospray, many peptides form [M+2H]2+ abundantly, and doubly charged precursors fragment into informative b and y ions spread across the spectrum.
- What is an immonium ion?
- A small ion formed from a single residue, its mass less carbon monoxide plus a proton. Immonium ions in the low-mass region indicate which amino acids are present, though not their order.
- Does a matching fragment spectrum prove the material is pure?
- No. It confirms the sequence of the ion that was selected. Purity is measured by chromatography.
Sources & Further Reading
- Roepstorff P, Fohlman J. Letter to the editors: proposal for a common nomenclature for sequence ions in mass spectra of peptides. Biomedical Mass Spectrometry 1984;11(11):601. doi:10.1002/bms.1200111109
- Biemann K. Contributions of mass spectrometry to peptide and protein structure. Biomedical and Environmental Mass Spectrometry 1988;16(1–12):99–111. doi:10.1002/bms.1200160119
- Steen H, Mann M. The ABC's (and XYZ's) of peptide sequencing. Nature Reviews Molecular Cell Biology 2004;5(9):699–711. doi:10.1038/nrm1468
- Paizs B, Suhai S. Fragmentation pathways of protonated peptides. Mass Spectrometry Reviews 2005;24(4):508–548. doi:10.1002/mas.20024
- Syka JE, Coon JJ, Schroeder MJ, Shabanowitz J, Hunt DF. Peptide and protein sequence analysis by electron transfer dissociation mass spectrometry. Proceedings of the National Academy of Sciences of the United States of America 2004;101(26):9528–9533. doi:10.1073/pnas.0402700101
See also
- HPLC purity analysis of peptides
- Mass spectrometry for peptide identity
- How to read a certificate of analysis
- Net peptide content and purity: two different numbers
- Peptide salt forms: trifluoroacetate, acetate and hydrochloride
- Amino acid analysis of peptides
- How synthetic peptides are made
- Endotoxin in peptide research materials
- Peptide stability and storage
- Chiral analysis: finding D-amino acid impurities
- Water content of peptides: Karl Fischer titration
- Related substances on a certificate of analysis
- The right mass does not prove the right peptide — myth versus measurement
- Tandem mass spectrometry — glossary
- Mass spectrometry — glossary
- Disulfide bond — glossary
Cite this page
Text
Cosmic Peptides. (2026, October 25). Sequencing peptides by tandem mass spectrometry. Cosmic Peptides Research Library. Retrieved October 26, 2026, from https://cosmicpeptides.com/research/methods/tandem-ms-sequencing
BibTeX
@misc{cosmicpeptides_research_methods_tandem_ms_sequencing,
author = {{Cosmic Peptides}},
title = {Sequencing peptides by tandem mass spectrometry},
howpublished = {Cosmic Peptides Research Library},
year = {2026},
month = oct,
url = {https://cosmicpeptides.com/research/methods/tandem-ms-sequencing},
urldate = {2026-10-26},
note = {Last reviewed 2026-10-25}
} RIS
TY - ELEC AU - Cosmic Peptides TI - Sequencing peptides by tandem mass spectrometry T2 - Cosmic Peptides Research Library PB - Cosmic Peptides PY - 2026 DA - 2026/10/25 UR - https://cosmicpeptides.com/research/methods/tandem-ms-sequencing Y2 - 2026/10/26 LA - en ER -
Cite the date the page was last reviewed, October 25, 2026, as its version. Reference managers such as Zotero can save this page directly.
Changes to this page
- Published Five more methods articles
Peptide stability and storage, chiral analysis, tandem mass spectrometry sequencing, water content by Karl Fischer titration, and related substances on a certificate of analysis; with three new glossary terms.
Changes of substance are listed here; wording changes are not. Every correction is also on the corrections log.
For in vitro laboratory research only. Not for human or veterinary use. This library describes chemistry, analytical methods and published laboratory research; it gives no guidance on use in people or animals.